2: DH-2000 Specifications
000-10000-025-02-1209 7
Rear Panel
Component Description
Variable Resistor
Use a screwdriver to adjust the intensity of the halogen lamp to optimize the
intensity between deuterium and halogen light in UV-VIS applications.
DH-2000-(S)-(FHS)-(DUV)-(BAL) Voltage Optical Power
Halogen Adjustment 5V - 12 V 10% - 100%
Power Terminal Input
Connect power cable to provide voltage to DH-2000:
Note: Only connect the power cable to the lamp when the Main Power Switch is in
the OFF position.
Main Power Switch
Turn on to supply power to the DH-2000. The Power LED illuminates when this
switch is in the On position.
Fuse
Contains the fuse to protect the unit against overload:
European Fuse Type: Miniature fuse 5 x 20 mm, 1 Amp slow blow
USA Fuse Type: Miniature fuse 5 x 20 mm, 2.5 Amp slow blow
Cooling Fan Cools the interior of the DH-2000. Do not obstruct.
Type Sign
Information about:
– Type – Version
– Order No. – Serial No.
– Main connection – Max. Ambient. Temperature
Warnings – CE-Marking
15-pin TTL Connector
(not shown)
On TTL Connector-equipped DH-2000 Series lamps, this connector allows for
external TTL control of lamp shuttering (-S and _BAL models only) and individual
lamp operation (TTL model only). See Pinout information.
Main Power Switch
Fuse
Cooling Fan
Type Sign
Variable Resistor