2: Installation
If you do not have web access, you can retrieve this document from the Software and Technical
Resources CD included with your spectrometer shipment.
NIR Controls
There are a number of controls for the NIR Spectrometer that are available in through the spectrometer
operating software. The software differs as to where these controls are located.
SpectraSuite NIR Spectrometer Features
Select Spectrometer | Spectrometer Features, then select the TEC tab to display the NIR Spectrometer
controls.
Control Description
Enable Thermo-
Electric Control
Check this box to enable (recommended setting for laboratory conditions) or
uncheck to disable the thermoelectric cooler.
Detector Set Point Enter the target temperature of the NIR Spectrometer detector as follows:
NIR512: Minimum is –10 ºC, Recommended is –5 ºC
NIR256: Minimum is –15 ºC, Recommended is –10 ºC
The TEC and fans will operate until the detector reaches this temperature.
Detector
temperature
Displays the current temperature of the detector in the NIR Spectrometer.