2: Installation
Control Description
Enable Periodic
Update
Check this box to enable periodic update of the NIR detector temperature
controls.
Update Interval If you enabled periodic updates, enter the interval period (in seconds).
Show in Status Bar Check this box to display the detector temperature in the status bar.
NIR Toolbar in OOIBase32
The NIR toolbar in OOIBase32 allows you to control a variety of options on the NIR Spectrometer. These
options are also applicable to the NIR256.
Note
If the NIR512 Toolbar in your version of OOIBase32 does not have the options listed
below, you should upgrade your version of OOIBase32 to the latest release.
Visit
http://www.oceanoptics.com/technical/softwaredownloads.asp to obtain the
latest software version.
Control Description
TEC On Check this box to enable (recommended setting for laboratory conditions) or
uncheck to disable the thermoelectric cooler.
Det. Temp. Set Point Enter the target temperature of the NIR Spectrometer detector as follows:
NIR512: Minimum is –10 ºC, Recommended is –5 ºC
NIR256: Minimum is –15 ºC, Recommended is –10 ºC
The TEC and fans will operate until the detector reaches this temperature.
Apply Click this button to apply the changes made in this toolbar to the NIR
Spectrometer.
Current Det. Temp. Displays the current temperature of the detector in the NIR Spectrometer.
197-00000-512-02-0707 9